YS6800 CT Contrast Injector – Technical Specifications
|
|
|
Model
|
YS6800
|
|
|
Main Controller Power Supply
|
AC 198–242 V, Frequency 50 ±1 Hz, 980 VA
|
|
|
Number of Channels
|
Three-channel
|
|
*
|
Maximum Injection Pressure
|
330psi
|
|
|
Injection Flow Rate
|
0.1–10 mL/s, increment 0.1 mL/s
|
|
*
|
Dual-flow Injection
|
Supports proportional mixing of contrast media and saline; mixing ratio 5%–95%
|
|
|
Injection Phases
|
Supports up to 12 injection phases
|
|
*
|
Alpha Assistant
(Intelligent Assistant)
|
Supported; fully automatic connection of inner and outer tubing, automatic air removal and line priming to injection-ready status; detects outer tubing connection to prevent misoperation
|
|
*
|
Overpressure Protection
|
User-selectable: pressure reduction followed by stop, or immediate stop upon overpressure
|
|
|
Air Detection Alarm
|
Supported
|
|
|
Ultrasonic Air Sensors
|
5
|
|
|
Automatic Left/Right Channel Switching
|
Supported
|
|
|
Automatic Inner Tube Up/Down Movement
|
Supported
|
|
|
Main Operation Screen
|
15-inch touchscreen
|
|
|
On-body Operation Screen
|
7-inch touchscreen
|
|
*
|
Splash-proof On-body Screen
|
Side-mounted design to prevent contrast or saline dripping and interfering with operation
|
|
|
Injection Program Storage/Recall
|
Stores up to 2,000 programmable injection protocols
|
|
*
|
Remaining Volume Display
|
Real-time display of remaining contrast volume on main or on-body screen
|
|
*
|
Automatic Remaining Volume Refresh
|
Automatic of remaining volume when bottle is changed mid-procedure
|
|
|
Injection / Venting Method
|
Direct ion of contrast bottle; no syringe, no bottle opening or aspiration required; automatic air removal
|
|
*
|
Magnetic Bottle Holder
|
Magnetic holder with 90° rotation for easier bottle removal; no latch operation required
|
|
*
|
Aluminum Alloy Heating Sleeve
|
Optional heating sleeve; easier bottle installation/removal; no need to open the sleeve
|
|
*
|
Data Transmission
|
Wired or wireless (optional)
|
|
*
|
KVO (Keep Vein Open)
|
Supported
|
|
|
Injection Delay
|
0-1000s
|
|
*
|
Inter-phase Delay
|
0-1000s
|
|
*
|
Timer
|
Supports start-triggered timing and injection-end-triggered timing
|
|
|
Test Injection via On-body Touchscreen
|
Supported; independent programming
|
|
*
|
One-touch Automatic Preparation
|
Intelligent one-button preparation to injection-ready status; no complex operation memorization required
|
|
*
|
Medication Change Detection
|
Three-channel sensing to detect bottle replacement by operator
|
|
|
Information Workstation
|
Not supported
|